|
|
||||||||
1 Dept of Respiratory Medicine & Allergy, Guy's, King's & St. Thomas' School of Medicine, King's College Hospital, London, UK. 2 Dept of Respiratory Diseases, University of Parma, Italy. 3 Respiratory Muscle Laboratory, Royal Brompton Hospital, London, UK
CORRESPONDENCE: A. Chetta, Istituto di Malattie Respiratoire, Azienda Ospedaliera e Università di Parma, Viale G. Rasori 10, Parma, Italy. Fax: 39 0521292615
Keywords: expiratory muscle strength, respiratory muscles, respiratory pressures, whistle mouth pressure test
Received: June 19, 2000
Accepted November 13, 2000
| Abstract |
|---|
|
|
|---|
To obtain a maximal whistle, subjects (27 healthy volunteers and 10 patients with amyotrophic lateral sclerosis) were asked to perform a short, sharp blow as hard as possible, from total lung capacity, through a reversed paediatric inhaler whistle, connected to a flange-type mouthpiece.
In both healthy subjects and patients, whistle mouth pressure (Pmo,W) was closely related to the pressure measured in the oesophagus and stomach during the same manoeuvre. In healthy subjects, Pmo,W and PE,max correlated with wide limits of agreement, although Pmo,W values were significantly higher than PE,max (131±31 cmH2O versus 101±27 cmH2O, p<0.0001). In patients, it was also found that Pmo,W and PE,max values were strongly related (r=0.937, p<0.0001). In healthy subjects, the intraclass correlation coefficient and the variation coefficient for Pmo,W repeated measurements were respectively 0.88 and 7.0%. However Pmo,W and PE,max were always smaller than the gastric pressure generated by a maximal cough.
It is concluded that mouth whistle pressure, a noninvasive, reproducible and simple test, provides a reliable measure of expiratory muscle strength in healthy subjects that is acceptable to patients and can be used in a complementary fashion to maximal static expiratory effort.
Chest infection is a cause of serious morbidity and mortality in patients with respiratory and neuromuscular disease. Cough is considered to protect against chest infection and it has recently been shown that expiratory muscle strength is an important determinant of an effective cough 1. Techniques to evaluate expiratory muscle strength are therefore of interest.
Mouth pressures during maximal static expiratory (PE,max) efforts are widely used as a noninvasive test to assess expiratory muscle strength 2. When high values are found, muscle weakness is excluded 3. However, the normal range for PE,max is wide 2, 4, reflecting both the biological variability of respiratory muscle strength and also the difficulty some subjects have in performing the manoeuvre maximally. In the assessment of inspiratory muscle strength it has been shown that the use of a manoeuvre complementary to the static effort, such as the maximal sniff, can exclude inspiratory muscle weakness 5, 6. Indeed sniffing is a natural manoeuvre which is more easily performed than static efforts, and by measuring the upper airway pressure during a maximal sniff, inspiratory muscle strength may now be accurately measured noninvasively 7 in a variety of clinical settings 8.
It was therefore reasoned that an additional test of expiratory muscle strength that was noninvasive, using a natural, uncomplicated manoeuvre, could be clinically useful. The main muscles of expiration are the abdominal muscles and one approach to measuring their strength is to record gastric pressure during a maximal cough 1 or following magnetic nerve stimulation posteriorly at the level of the 10th thoracic intervertebral space 9. However, both these tests require passage of a gastric balloon catheter. The present authors hypothesized that a test using pressure measured at the mouth during a short, sharp and maximal expiration through a narrow aperture (a "whistle") would be a relatively natural manoeuvre that might be a useful additional noninvasive test of expiratory muscle strength. The main purpose of the study was to ascertain in healthy subjects, and in patients with respiratory muscle weakness, whether the measurement of mouth pressure during a whistle manoeuvre (Pmo,W) is reproducible, and whether mouth whistle was pressure closely reflects oesophageal and gastric pressures. In addition, the study aimed to compare mouth, oesophageal and gastric pressures during whistle and PE,max manoeuvres, as well as to compare whistle mouth pressure to gastric pressure, during a maximum cough.
| Methods |
|---|
|
|
|---|
Spirometry was measured by a horizontal bellows spirometer (Vitalograph Ltd, Buckingham, UK) and vital capacity (VC), forced expiratory volume in one second (FEV1) and FEV1/VC values were recorded. Predicted values of VC and FEV1 were obtained from regression equations by Quanjer et al. 12.
In order to obtain a maximal whistle, subjects were asked to perform a short, sharp blow as hard as possible from total lung capacity (TLC) through a reversed paediatric inhaler whistle (Astra Pharmaceuticals Ltd, Herts, UK) connected to a flange-type mouthpiece gripped between the teeth (fig. 1
). Subjects performed whistles without a noseclip. Additionally, healthy subjects were asked to perform maximum whistles while holding their cheeks. Whistles suitable for analysis had to present pressure tracings showing a sharp peak and a duration of <500 ms. This was usually achieved after 36 whistle manoeuvres. Healthy subjects were also asked to perform whistles of varying intensities. Repeatability of whistle mouth pressure (Pmo,W) in healthy subjects was assessed by repeating measurements on two separate days. The whistle device had a linear relationship between resistance and flow across the range 10100 L·min1. The mean (range) value for peak flow through the device for healthy subjects was 92 L·min1 (78109 L·min1). The resistance over this range was 29.344.4 cmH2O·L1·s.
|
Gastric (Pga), oesophageal (Poes) and transdiaphragmatic (Pdi) pressures were measured using a pair of commercially available latex balloon catheters (PK Morgan, Rainham, Kent, UK) 110 cm in length, passed through the nose and positioned in the stomach and oesophagus in the conventional manner 14. The oesophageal balloon contained 0.5 mL air, the gastric balloon 2 mL.
Pga during a maximum voluntary cough effort, was also measured in healthy subjects and patients. This manoeuvre was performed in a seated position without a noseclip. Invariably, subjects inhaled deeply before coughing, but no specific instructions were given regarding the magnitude of inspiration before cough. Repeated efforts were performed until no further increase in Pga was obtained (usually achieved after 36 coughs).
The whistle device, the mouth piece and the balloon catheters were connected to differential pressure transducers (Validyne MP45-1, Validyne, Northridge, CA, USA), carrier amplifiers (PK Morgan), a 12 bit NB-MIO-16 analogue-digital board (National Instruments, Austin, TX, USA) and a Macintosh Quadra Centris 650 personal computer (Apple Computer Inc., Cupertino, CA, USA) running LabviewTM software (National Instruments). Pdi was obtained on-line, by subtraction of Poes from Pga. A minimum sampling frequency of 100 Hz was used. Peak pressures were taken as those measured from the baseline at relaxed end-expiratory lung volume, at the peak pressure, obtained.
Healthy subjects and patients were studied in the seated position and pressures were displayed on a computer screen in front of the subject to provide visual feedback 15. Subjects were strongly encouraged to make maximal efforts.
Data are presented as mean±sd. VC and FEV1 values were expressed as per cent of predicted value, FEV1/VC is expressed as per cent. Differences in numerical data were examined by means of paired t-tests. The agreement between measures was assessed by the method of differences against the means according to Bland and Altman 16. The relationship between measures was assessed by Pearson's correlation coefficient (r) and linear regression analysis. The repeatability of measures was expressed as intraclass correlation coefficient (rI) 17 and assessed by calculating the coefficient of variation. A p-value <0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
In the 27 healthy subjects, Pmo,W and PE,max values were 131±31 cmH2O (range 77202 cmH2O) and 101± 27 cmH2O (range 44155 cmH2O), respectively (p< 0.0001). The 95% confidence intervals of the mean were 119143 cmH2O for Pmo,W and 90111 cmH2O for PE,max. In the same group of subjects, Pmo,W did not differ from Pmo,W obtained while holding cheeks (139±35 cmH2O, range 72217 cmH2O). The ratio Pmo,W/PE,max was 1.3 (range 0.92.1) and in 24 out of 27 subjects, was >1. There was a significant correlation between Pmo,W and PE,max (r=0.674, p=0.0001) (fig. 2
). The bias between Pmo,W and PE,max was 30 cmH2O, with Pmo,W tending to be numerically greater than PE,max. The limits of agreement ranged from -1777 cmH2O (fig. 2
). The rI and the variation coefficient for Pmo,W repeated measurements were 0.888 and 7.0%, respectively. In the same group of subjects, rI and variation coefficient for PE,max were 0.790 and 10.2%, respectively.
|
|
|
|
|
The mean values of Pmo,W and PE,max were 69± 32 cmH2O (range 17128 cmH2O) and 62±30 cmH2O (range 11116 cmH2O), respectively. The 95% confidence intervals of the mean were 4692 cmH2O for Pmo,W and 4184 cmH2O for PE,max. The ratio Pmo,W/PE,max was 1.17 (0.871.55). There was a significant correlation between Pmo,W and PE,max (r=0.937, p< 0.0001). The bias between Pmo,W and PE,max was 7 cmH2O and the limits of agreement ranged -1529 cmH2O (fig. 3
).
The mean Poes, Pga, and Pdi values during both whistle and PE,max manoeuvres were 66±32 cmH2O and 63±29 cmH2O, 74±37 cmH2O and 56±33 cmH2O (p<0.02), and 8±7 cmH2O and -7±10 cmH2O (p<0.002), respectively. The ratios of Pmo,W/Poes,W and Pmo,W/Pga,W were 1.08 (range 0.821.31) and 1.17 (range 0.871.55). Biases between Pmo,W and Poes,W and between Pmo,W and Pga,W were 3 cmH2O and -5 cmH2O, respectively and limits of agreement ranged -1016 cmH2O and -199 cmH2O (fig. 3
). There was a significant correlation between Pmo,W and Poes,W (r=0.980, p<0.0001) and between Pmo,W and Pga,W values (r=0.990, p<0.0001). Regression analysis showed that Pmo,W=-5.33+1.1Pga,W and Pmo,W=-1.78+0.9Poes,W.
The mean value of cough Pga was 85±39 cmH2O (p<0.05 versus PE,max) and the 95% confidence intervals of the mean were 57112 cmH2O. Ratios of Pmo,W/cough Pga and PE,max/cough Pga were 0.88 (range 0.501.39) and 0.78 (0.441.36), respectively. Biases between Pmo,W and cough Pga and between PE,max and cough Pga were -16 cmH2O and -23 cmH2O, respectively, and limits of agreement ranged -7543 cmH2O and -9145 cmH2O, respectively (fig. 4
). There was a significant correlation between Pmo,W and cough Pga (r=0.67, p<0.05).
| Discussion |
|---|
|
|
|---|
Critique of the method
Do mouth measurements reflect expiratory muscle strength?
The aim of the study was to develop a test which reflected abdominal muscle strength, since these muscles are the main muscles of active expiration and are an important determinant of cough function. In the present study, Pga was almost invariably greater than Poes during both Pmo,W and PE,max indicating that the abdominal muscles are the driving force in both manoeuvres. In healthy subjects, for whom Poes, Pga and Pdi were measured during both manoeuvres, Pga was similar for both manoeuvres, though Poes was different, resulting in higher Pdi values with PE,max compared to Pmo,W. This suggests that Pmo,W may be a more accurate reflection of abdominal muscle strength than PE,max.
However, for the group as a whole, Pga was incompletely transmitted from the abdomen to the chest; thus Poes was 79% and 68% of Pga during whistle and PE,max manoeuvres, respectively. The reduced transmission of gastric pressure could be the result of diaphragm activation and the "inspiratory" action of the abdominal muscle on the lower rib cage 19, 20. As further confirmation that Pmo,W closely reflects both Poes,W and Pga,W, a strong correlation between Pmo,W and Pga,W and between Pmo,W and Poes,W was found in healthy subjects over a wide range of pressures obtained during whistles of different intensity. This predicts that the Pmo,W test will remain valid in patients with expiratory muscle weakness, a prediction supported by the data from ALS patients shown in figure 3
. Most importantly, it was found that, in patients with ALS of varying severity, Pmo,W values strongly correlated with Pga,W.
Dynamic nature of the manoeuvre
Unlike the PE,max test, which is based on a static manoeuvre, the whistle test is based on a dynamic one. With dynamic tests, concern can arise because airflow can render the manoeuvre not truly isovolumic. For the sniff this tendency is minimized because the nose acts as a Starling resistor, so that nasal flow is reduced, largely independent of driving pressure 21. For the whistle, our in vitro assessment showed that flow increased with driving pressure and this might have resulted in Pmo,W being lower than PE,max; this was not so, indicating that the volume expired before peak pressure is not sufficient to importantly detract from the value of the test. This is because the whistle orifice acts as a resistor, substantially reducing expiratory flow. In this regard, the mean value we measured in healthy subjects for peak flow through the device, was 92 L·min1, approximately 68 times lower than their peak expiratory flow rate.
Recoil pressures
PE,max is normally measured at TLC because subjects find it easier to maximize expiratory efforts at high volumes. In this study the whistle test manoeuvre was also performed at TLC and, at this lung volume, the pressure measured reflected both the pressure developed by the expiratory muscles and the passive elastic recoil pressure of the respiratory system. Specifically, at TLC the passive elastic recoil pressure can add up to 40 cmH2O, equal to the pressure generated by the expiratory muscles 22. However the whistle is a dynamic manoeuvre and recent data obtained using the peak expiratory flow 23 suggest the possibility that the recoil force generated by the chest wall might be greater in this situation than after the static manoeuvre. Whilst this might contribute to Pmo,W being higher than PE,max, it does not explain the wide limits of agreement and therefore does not alter our conclusion that Pmo,W could be of value as a complementary test of expiratory muscle strength.
Facial muscles
Conceptually, the facial muscles might contribute to Pmo,W, although the orifice of the whistle should serve the same function as the mouth leak proposed by Rinqvist 13. Consistent with this, the present study found no significant differences between Pmo,W measurements taken with or without cheek support and it is therefore concluded that this is not a matter of practical clinical importance.
Significance of the findings
In healthy subjects, it was found that Pmo,W and PE,max values correlated with wide limits of agreement, although Pmo,W values were significantly higher than PE,max and in all but three subjects the Pmo,W/PE,max ratio was >1. In ALS patients, it was also found that Pmo,W and PE,max values were closely related. Pmo,W mean values were numerically, but not statistically significantly, higher than those of PE,max, and the Pmo,W/PE,max ratio was >1 in seven of the 10 patients.
Although the PE,max manoeuvre is not complicated, it is not one with which many subjects will be familiar on a daily basis and requires full cooperation from subjects. In contrast, the whistle manoeuvre may be more natural and has the advantage of audible feedback. Low PE,max values can occur even in healthy subjects, perhaps due to a lack of motivation or poor technique and thus may not indicate reduced expiratory muscle strength. Predicted PE,max values are readily available for adults, the elderly, and children 2, 24 but the normal ranges are wide, making it difficult to accurately identify weakness. In the present study, the whistle manoeuvre was easily performed, with minimal instruction, and was well accepted both by healthy subjects and ALS patients. The high mouth pressures achieved and the relatively high lower limits suggest that the Pmo,W could be a clinically useful, additional method for assessing expiratory muscle strength. The day to day reproducibility of Pmo,W was good, suggesting the test could also be useful for serial measurements.
The ALS patients had variable severity of disease; some had severe respiratory muscle weakness whereas others had normal strength. In the patients, as in healthy subjects, the whistle manoeuvre was found to generate gastric, oesophageal and mouth pressures that were higher than those generated at PE,max. In particular, Pga values during whistle were significantly higher than those during PE,max, indicating that the whistle manoeuvre could better recruit abdominal muscle contraction than PE,max. Because of the small sample size, bulbar and nonbulbar patients were not specifically compared; clearly a theoretical concern could be that bulbar patients do not achieve satisfactory transmission of pleural pressure to the mouth during a whistle.
Interestingly, in ALS patients, Poes values were less than Pga values during whistle, but were higher than Pga values during PE,max. This might suggest that the rib cage expiratory muscles play a different role during the two manoeuvres with a greater involvement during the PE,max. Alternatively, one could hypothesize that the process of ALS leads to compensatory hypertrophy or recruitment of muscles not yet affected by the disease as suggested by Attali et al. 25.
Pmo,W and PE,max values were significantly lower than corresponding cough Pga values, both in healthy subjects and in ALS patients. Additionally, gastric pressure during whistle and PE,max manoeuvres was numerically lower than cough Pga. The present data confirm the results of a previous paper 1 that showed that cough is the most physiological expiratory manoeuvre and that cough Pga is the gold standard for measuring pressure caused by abdominal muscle contraction. However, it was also observed that the bias and the limits of agreement between Pmo,W and cough Pga were respectively smaller and narrower than those between PE,max and cough Pga, both in patients and healthy subjects. Additionally, in ALS patients, Pmo,W values did not significantly differ and were strictly related to cough Pga. Lastly, the shape of the pressure tracings recorded during the whistle manoeuvre were very similar to those during cough (fig. 5
). Considered together, these findings suggest that whistle is a manoeuvre physiologically closer to cough than PE,max.
In conclusion, this study indicated that, in both healthy subjects and in amyotrophic lateral sclerosis patients, mouth whistle pressure is an accurate reflection of expiratory muscle strength. The limits of agreement between mouth whistle pressure and maximum expiratory pressure were wide, suggesting a role for mouth whistle pressure as an additional method for the evaluation of expiratory muscle strength, for use in both in the physiology laboratory and in clinical settings.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Chatwin, E. Ross, N. Hart, A.H. Nickol, M.I. Polkey, and A.K. Simonds Cough augmentation with mechanical insufflation/exsufflation in patients with neuromuscular weakness Eur. Respir. J., March 1, 2003; 21(3): 502 - 508. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Hart, D. Cramer, S.P. Ward, A.H. Nickol, J. Moxham, M.I. Polkey, and N.B. Pride Effect of pattern and severity of respiratory muscle weakness on carbon monoxide gas transfer and lung volumes Eur. Respir. J., October 1, 2002; 20(4): 996 - 1002. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |