Copyright ©ERS Journals Ltd 2001 Airway resistance measured by the interrupter technique: expiration or inspiration, mean or median?Queen Elizabeth Children's Service, Royal London Hospital, Whitechapel, London CORRESPONDENCE: S. McKenzie, Queen Elizabeth Children's Service, Royal London Hospital, Whitechapel, London, E1 1BB. Fax: 44 1713777325 Keywords: airway resistance, interrupter resistance
Received: December 12, 1999
This work was supported by GlaxoWellcome and Queen Elizabeth Hospital for Children Trustees. Abstract
The measurement of airway resistance by the interrupter technique (Rint) needs standardization. Should measurements be made be during the expiratory or inspiratory phase of tidal breathing? In reported studies, the measurement of Rint has been calculated as the median or mean of a small number of values, is there an important difference?
Subjects were 2.55.0 yrs (median 4.0 yrs) who had previous respiratory symptoms. The Rint in expiration (RintE) and inspiration (RintI) pre and postsalbutamol, the coefficient of variation (CV) of values contributing to measurements, and bronchodilator responsiveness (BDR) in both phases were compared. Measurements using median and mean were compared.
RintE was higher than RintI by 4% (p<0.01). The CV of values making up RintE and RintI, and BDR measured in expiration and inspiration were similar. The median difference between means and medians of values making up measurements was 0.6% (range 611%).
RintE has been shown to be consistently greater then RintI but the difference in this study is small. It is suggested that one or the other is chosen as the standard. In the present data the mean of a set of values contributing to a measurement was not significantly different from the median. However, the use of the median has been recommended since it is less affected by possible outlying values such as might be included by fully automated equipment.
Early Rint investigations were inconsistent in the use of the expiratory or inspiratory phase of tidal breathing. More recently, Phagoo 3, in his work on the analysis of Pmo(t) transients, on which some commercial devices' algorithms are based, used the expiratory phase of tidal breathing to try to obtain a signal with minimum interference from muscular activity 3. However, others have used inspiration 5, 6 because of concern about variations in glottic opening which are more likely to have an effect on the measurement during expiration 7. Some workers have observed that, during expiration, children sometimes blow in anticipation of the trigger. If there are more technical or physiological difficulties when measuring Rint in expiration (RintE) or inspiration (RintI) these may be reflected in a greater number of unacceptable Pmo(t) transients and may be reflected in a larger coefficient of variation (CV) of the constituent values. Typically 510 Rint values contribute to a measurement, in studies in young children. It has been suggested that the median of these values should be the measurement because outlying values will affect the average. Most published studies have used the mean 1, 3, 5, 6 and so it is important to know how mean and median compare. The purpose of this study was to measure the difference between RintE and RintI and to compare the mean and median measurements. Methods
Interrupter resistance measurements The subject came off the mouthpiece for 35 breaths between values and for 30 s between RintE and RintI. Measurements were repeated 15 min after inhalation of 400 µg salbutamol via a spacer device. RintE and RintI were obtained with the values double blind to the technician and patient until completion of the test. For baseline measurements, the total number of interruptions was recorded, in each mode, so that the number of interruptions for six acceptable values could be counted. The time to obtain the baseline measurement, six acceptable values, was recorded to the nearest one-quarter of a minute, as this was the shortest time which could be accurately measured.
Subjects
Statistical analyses Results
Measurements of RintE and RintI are shown in figure 1
Baseline measurements of flow during expiration and inspiration are shown in figure 2
There was no difference in the measurement of bronchodilator responsiveness (BDR), the ratio of Rint beforebronchodilator to Rint after bronchodilator, in the expiratory or inspiratory phase (RintEbaseline:RintEpostsalbutamol=1.27; RintIbaseline:RintIpostsalbutamol=1.31, p=0.45). The CV of RintE measurements did not differ from RintI measurements (mean baseline CV E=15.5% and I=16.5%, 95% CI for difference=3.51.4; mean postsalbutamol CV RintE 18.1% and RintI 18.2%, 95% CI=3.02.9%). The number of interruptions required for six acceptable values did not differ in the expiratory and inspiratory phases (median 6 for both phases; range 610). There was no difference in the time taken to obtain a set of six values (median 4.5 min; range 46.75 min). The mean:median ratios for a random selection of 100 sets of values of 69 per set (each subject had two baseline and two postsalbutamol measurements made) were not normally distributed. The median ratio was 1.006 (range 0.941.11). This implies that for the present data the mean was 6%11% from the median. Discussion In this study, a geometric mean difference of 4% between Rint measurements in the expiratory and inspiratory phases of tidal breathing has been demonstrated. There are slightly higher differences at the lowest measurements made. The measurements in this group of preschool children are similar to those published 10. Although flows were significantly lower in expiration than in inspiration, no correlation between flow and corresponding Rint measurements and no correlation between change in flow and change in Rint was shown. It is possible that in airways where there is laminar flow, lateral negative pressure on the airway wall may promote airway narrowing. This, the Bernoulli effect 11 would be expected both in inspiration and in expiration. Another reason for airway narrowing in expiration may be increased airway compliance as intrapleural pressure becomes less negative. Resistances throughout expiration have been demonstrated to be higher than in inspiration in a younger group of recurrently wheezy subjects 12. The subjects in the present study included children who had previously been wheezy. It is not known whether there is change in glottic diameter during tidal breathing in children, but, if anything, narrowing would be expected in inspiration rather than in expiration due to apposition of supraglottic soft tissues. Blowing or sucking in anticipation of the trigger in expiration or inspiration was not observed. The device used has a random trigger that occludes after varying numbers of breathing cycles. This should overcome this potential problem. The system, as currently configured, does not allow for the determination of where interruption occurs in relation to volume. Measurements in the laboratory in similar subjects using a newer system (MicroRint, Micro Medical) which displays flow against time, indicate that interruption occurs at the start of expiration and at the start of inspiration. If these timings correspond to high and low lung volumes respectively, then RintE would be expected to be lower than RintI, the opposite of the present findings. Differences in lung volume at the time of interruption in the two phases would not, then, explain the differences in resistance. The repeatability of the measurement in the laboratory, (2 sd of the mean ratio of two measurements 30 s apart) 1 is ±16%. The difference between measurements in expiration and inspiration is small by comparison. Only two other studies, in older children, have systematically investigated RintE and RintI 2, 8 and have shown RintE to be greater than RintI. Differences are of the order of 20% in both, much higher than has been demonstrated. There is no explanation for this other than noting that the age and health status of subjects, and methods are different. There are various methods used to describe BDR. The ratio of baseline to postbronchodilator was chosen. No difference in BDR measured in the expiratory or inspiratory phases was found, as others have shown 8. A within-subject CV of the set of values making up a measurement has been considered to be acceptable if it is <20% 2. The CV of the values is on average <20%. CV measured in expiration or in inspiration is similar. There was no difference between expiration and inspiration in the number of interruptions needed to obtain acceptable values, nor the time to do the test. It is recommended that the median of the six or so values should be used as the measurement, as this is less affected than the mean by outlying values. Such values cannot be avoided with fully automated equipment in which transients cannot easily be examined for acceptability. Some commercial devices collect values which are considered by the machine's software to be acceptable. Inexperienced operators may overlook outlying values which could have resulted from poor technique, such as moving the head or blowing the cheeks out. It has been shown that there was very little difference between mean and median values in the data, so that studies which have used the mean are still valid 1, 3, 5, 6. In conclusion, this study has shown that interrupter resistance in expiration is 4% higher than in inspiration, a value much smaller than the repeatability of the measurement. The mean and median of values which contribute to a measurement using this method differ very little. The authors agree with Carter 4 that there is a need for standardization of all aspects of interrupter resistance measurement.