|
|
||||||||
a Mt. Sinai Hospital, University of Toronto, Toronto, Canada, b Memorial Hospital of Rhode Island and c Rhode Island Hospital and Brown University School of Medicine, Providence, RI, USA
CORRESPONDENCE: N.S. Hill, Division of Pulmonary, Sleep, and Critical Care Medicine, Rhode Island Hospital, 593 Eddy Street, Providence, RI 02903, USA. Fax: 1 4014446665
Keywords: mask ventilation, mechanical ventilators, noninvasive ventilation, portable ventilators
Received: September 30, 1999
Accepted August 21, 2000
| Abstract |
|---|
|
|
|---|
Using a test lung model, compensatory capabilities of the ventilators were tested for smaller and larger leaks using the assist/control or timed modes. Back-up rate was 20·min1, inspiratory pressure was 18 cmH2O, and expiratory pressure was 5 cmH2O.
It was found that even in the absence of air leaking, delivered tidal volume differed substantially between the ventilators during use of pressure-targeted modes, depending on inspiratory flows, inaccuracies in set versus delivered pressures, and inspiratory duration. Also during pressure-targeted ventilation, increasing the tI/ttot up to, but not beyond, 0.5 improved compensation by lengthening inspiratory duration, whereas use of a sensitive flow trigger setting tended to cause autocycling during leaking, interfering with compensation. Leaking interfered with cycling of the BiPAP S/T, inverting the I:E ratio, shortening expiratory time, and reducing delivered tidal volume. Volume-targeted modes achieved limited compensation for small air leaks, but compensated poorly for large leaks.
To conclude, leak-compensating capabilities differ markedly between ventilators but pressure-targeted ventilators are preferred for noninvasive positive pressure ventilation in patients with substantial air leaking. Adequate inspiratory flows and durations should be used, triggering sensitivity should be adjusted to prevent autocycling, and a mechanism should be available to limit inspiratory time and avoid I:E ratio inversion.
Noninvasive positive pressure ventilation (NPPV) is widely used to assist breathing in both acute and chronic forms of respiratory failure 1, 2. In contrast to invasive ventilation, NPPV uses an open circuit design that is inherently leaky. Although ventilation can often be assisted even in the presence of sizeable leaks 3, different ventilators and ventilator modes may be more or less capable of compensating for air leaks. This compensatory capability may be important in optimizing the success of NPPV, but few studies have examined the effectiveness of different ventilators in delivering ventilation in the face of air leaks.
Air leaks during NPPV consist of either mask leaks between the skin and mask, mouth leaks with nasal ventilation, or nose leaks with mouthpiece ventilation. Some air may also escape via the oesophagus, but because of the relatively high impedance posed by the lower oesophageal sphincter 4, this route is likely to be minor in comparison to the others. Studies on patients using nocturnal volume- or pressure-limited ventilation have found that air leaks occur during most of sleep 3, 5. These studies show that ventilation and oxygenation are adequately supported in most patients despite the presence of leaks, but that large leaks may interfere with ventilator cycling, compromise minute ventilation, and cause sleep fragmentation due to leak-associated arousals 3, 5.
Simple bedside interventions may alleviate air leaking. Mask leaks can be reduced by ensuring proper mask fit and by using optimal headstrap tension. The temptation to merely tighten the straps to reduce air leaking must be resisted, because this may reduce patient comfort and tolerance of NPPV, and promote the development of nasal bridge ulcers 6. Air leaking through the mouth that occurs during nasal ventilation can be reduced by encouraging patients to keep their mouths closed, using chin straps, or switching to a mouthpiece or an oronasal mask 6. However, air leaks often persist despite these interventions.
For these reasons, ventilators that are designed to compensate for air leaks are desirable for noninvasive ventilation. Using a lung model testing system, the leak compensating abilities of six positive pressure ventilators commonly used to administer NPPV, including pressure-targeted and volume-targeted devices were evaluated and compared. It was hypothesized that ventilator characteristics such as mode of ventilation, peak inspiratory flow capabilities, triggering sensitivities, and criteria for termination of inspiration as well as lung characteristics such as resistance and compliance, would determine the effectiveness of leak compensation.
| Methods |
|---|
|
|
|---|
All ventilators were tested in controlled modes using a back-up rate.
The BiPAP ventilator was evaluated in both the spontaneous/timed (S/T)
and timed (T) modes, and the Quantum, 335, and PLV 102 were evaluated
in the assist/control mode. For the O'NYX and Siemens ventilators,
both pressure-targeted (assist controlled pressure ventilation (ACPV)
and pressure control ventilation (PCV), respectively) and volume-targeted (assist
control ventilation, ACV) modes were evaluated. Important differences
between the ventilators are listed in table 1
. These include differences in inspiratory termination (cycling)
criteria between the ventilators. The Quantum 335 and Siemens (for both
modes) use a timer set by the I:E ratio when inspirations are triggered
by the timed back-up rate, whereas BiPAP in the S/T mode cycles
in response to a decrease in inspiratory flow 7. For this reason, the timed (T) mode on the BiPAP
that cycles into expiration based on a preset timer was also tested. Ventilator
tubing used was that recommended by the manufacturer. The Whisper SwivelTM, a fixed exhalation valve, was used with the BiPAP ventilators. For
the O'NYX, Siemens and PLV 102, separate inspiratory and expiratory
circuits and exhalation valves were used as supplied by the manufacturer.
|
|
Experimental settings
For baseline comparisons, respiratory frequency was set at 20 breaths·min1 in all of the experiments to reflect commonly recommended breathing
rates for noninvasive ventilation 8, 9. IPAP/EPAP settings were 18 and 5 cmH2O respectively for the BiPAP (in both T and S/T modes),
335, and Quantum ventilators. Inspiratory and expiratory sensitivity settings
on the 335 were 2 and 3, respectively, with 1 being the most and 5 the least
sensitive settings 7. In order
to match the IPAP/EPAP settings, the O'NYX in the ACPV mode was
at a pressure support of 13 cmH2O and a PEEP of 5 cmH2O and the inspiratory trigger was set at 1.5 cmH2O (0
being the most and 3 the least sensitive settings). Inspiratory time/total
breath time (tI/ttot)
was set at 0.33, and test lung compliance at 0.1 L·cmH2O.
To test the effect of differences in inspiratory duration on delivered VT, tI/ttot was
varied (0.25, 0.33 or 0.50) for the BiPAP and PLV. The BiPAP T was
additionally tested at a tI/ttot
of 0.66 to determine the effect of inverted I
:
E
ratios on delivered VT. The influence of inspiratory trigger
sensitivity was tested by using the O'NYX at inspiratory sensitivities
of 0.5, 1.5 and 3.0 cmH2O. The effect of changes in respiratory
system compliance and resistance on tidal volumes delivered in the presence
of leak were evaluated using the 335 at test lung compliances of 0.1, 0.06
and 0.03 L·cmH2O. With test lung compliance set at
0.1 L·cmH2O, each of 3 resistance conditions (no
resistor, Rp 5 or Rp 20) was also tested. The effect of
varying leak on VT delivered by volume-targeted ventilation
was assessed using the PLV 102, O'NYX and Siemens at four different tidal
volume settings, with test lung compliance set at 0.1 L·cmH2O and tI/ttot at 0.33.
Statistical analysis
Six breaths were analysed and averaged for each experimental setting. All
values are reported as mean±sd. Repeated measures analysis
of variance was used to compare mean values for each ventilator at different
settings, and two way analysis of variance was used to compare mean values
between different ventilators. When significant differences were detected, post hoc analysis was performed using Tukey's exact test. Very little
variability occurred between breaths, so sd bars are not shown
in the figures to improve clarity of presentation.
| Results |
|---|
|
|
|---|
|
|
|
Effect of set inspiratory time/total breath time on delivered tidal volume
Among the "bilevel" devices that permit setting of tI/ttot, delivered VT increased
as tI/ttot was raised from 0.25,
to 0.33 and 0.5, both with small and large leaks (fig. 3
). At each setting, the Quantum preserved
leak VT better than the 335 or BiPAP T, related to a longer
inspiratory duration (table 2
).
During air leaking, the BiPAP in the S/T mode delivered a lower VT (fig. 4
).
This was related to the fact that in the S/T mode, the % IPAP control
on the BiPAP is inoperative and does not permit adjustments of tI/ttot. Inspiration is terminated by a decrease
in inspiratory flow up to a maximum duration of 3 s; therefore, tI/ttot increases during leak because
inspiratory flow fails to drop sufficiently to cycle the ventilator. tI/ttot averaged 0.47 in the absence
of leak, and increased to 0.8 with a small or large leak. To illustrate this
effect of excessive prolongation of tI/ttot, delivered VT using the BiPAP in the T mode
fell from 1.23 L to 0.83 L when tI/ttot was lengthened from 0.5 to 0.66, while other settings
remained unchanged.
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Even in the absence of leaks at comparable settings, the five pressure-targeted ventilators in the present study delivered quite different VTs. The factors accounting for these differences included inaccuracies in delivered versus set inspiratory and expiratory pressures and a longer inspiratory duration when the BiPAP was set in the S/T mode, allowing more time for lung inflation. Because of this property, the BiPAP in the S/T mode also exhibited the greatest proportional reduction in delivered VT during leaks. This occurred because the sustained high inspiratory flow during leaks prevented termination criteria for inspiration from being reached for up to 3 s, the default setting. At the set respiratory frequency of 20 breaths·min1, this led to an inversion of the I:E ratio to as high as 4:1. VT fell because of incomplete emptying of the lung due to shortening of the expiratory time.
In the presence of leaks, pressure-targeted modes manifested even larger
differences in delivered VT than without leaks. This was
related to the ability to increase inspiratory flow and sustain the target
inspiratory pressure as well as the duration (inspiratory time).
The ventilators that best compensated for leaks (especially large leaks
such as the Quantum) were able to increase inspiratory flow rate up to
three-fold and lengthened inspiratory duration slightly (table 2
). Excessive prolongation of inspiratory
time was counterproductive, however, as discussed above. In addition, by demonstrating
the reduction in inspiratory flow that occurs with the Siemens ventilator
in the face of large leaks (fig. 3
),
the results illustrate the importance of not only attaining a high inspiratory
flow rate during leaks, but also sustaining it for the duration of the inspiratory
phase.
In contrast to the pressure-targeted modes, ventilators delivering volume-targeted modes compensated less well for leaks because inspiratory flow and duration are either fixed (PLV 102) or increase slightly (O'NYX and Siemens). Accordingly, significant drops in delivered VT (at least 50%) during small leaks and more than 80% during large leaks were observed. In addition, prolonging tI/ttot from 0.25 to 0.5 did not increase delivered VT as it did with pressure-targeted modes. In fact, delivered VT fell slightly, presumably because of the greater inspiratory duration and resulting longer leak time. In the presence of a small leak, compensation was achieved by increasing set VT to 2 L as delivered VT was raised to 500, 900 and 1200 mL for the PLV 102, Siemens and O'NYX ventilators, respectively. Nevertheless, this strategy for leak compensation is less effective than using pressure-targeted ventilators. Thus, volume-targeted ventilators would not be the first choice for NPPV in patients with substantial air leaking.
Some of the pressure-targeted ventilators permit setting of the I:E ratio (the BiPAP T mode, 335, and Quantum). Prolonging the inspiratory duration by increasing the set tI/ttot from 0.25 to 0.33 or 0.5 improved leak compensation. However, this compensatory effect depends on the rate of lung filling and emptying and the absolute inspiratory duration. For instance, increasing the tI/ttot beyond 0.25 at a rate of 10 would probably have less effect than at a rate of 20 because the inspiratory duration at the lower rate would allow complete lung filling 13. On the other hand, prolonging the inspiratory time to the point of inverting the I:E ratio is counter-productive at a rate of 20, as exemplified by the drop in delivered VT when the tI/ttot was increased from 0.5 to 0.66 on the BiPAP in the T mode. Other potential clinical consequences of inversion of the I:E ratio include the need for patients to activate their expiratory muscles of respiration in order to cycle into EPAP, air trapping due to inadequate expiratory time, and patient-ventilator asynchrony that contributes to patient discomfort and NPPV intolerance. Thus, the capability of limiting tI/ttot to no more than 50% of the respiratory cycle time can help to prevent excessive shortening of expiratory time. In patients with severe COPD, a shorter tI/ttot may be desirable to enhance patient-ventilator synchrony. Previous studies have shown that relatively high inspiratory flow rates (and hence short inspiratory times) reduce work of breathing during pressure support breathing in these patients 14.
The present results also illustrate the potential adverse consequences of sensitive flow triggers during leaking. At the more sensitive inspiratory trigger settings on the O'NYX and Siemens ventilators, air leaking caused autocycling and marked reductions in delivered VT due to the shortening of inspiratory and expiratory times. Flow-triggered ventilators are susceptible to autocycling in the presence of a leak because the leak flow may be interpreted by the ventilator as the onset of inspiration. In a study evaluating three paediatric flow-triggered intensive care ventilators 15, the relative rate of autocycling by the three ventilators was determined by the sensitivity setting, and all of the ventilators autocycled less frequently at decreased sensitivity settings. The presented findings are consistent with these, with rapid respiratory rates occurring during both small and large leaks. These results indicate that monitoring should be performed during use of ventilators with adjustable inspiratory triggers, and trigger sensitivity should be set to avoid autocycling. It should also be noted that certain ventilators are resistant to autocycling by design. For instance, the BiPAP S/T that did not autocycle in our study, has a higher threshold for inspiratory triggering early compared to later during expiration 16.
As anticipated, increases in respiratory impedance caused by increases in test lung elastance or resistance resulted in reductions in VT delivered by the pressure-limited modes, but had little effect on VT delivered by the volume-targeted PLV 102. Thus, in the setting of increased airway resistance or respiratory system elastance, there are several choices available to the clinician. If leaks are minimal, volume-targeted ventilation can be used to assure delivered VT despite the increased impedance. However, because leaks are so common during NPPV, pressure-targeted ventilators may be preferred to compensate for the leaks as long as ventilator pressures or inspiratory times are adjusted to optimize delivered VT. However, the effectiveness of these latter compensatory actions is limited by the patient's tolerance of increased pressures during NPPV, and the increased pressures could exacerbate the leak 17.
A number of limitations should be borne in mind when interpreting the present results. First, a test lung was used in order to compare the ventilators under identical mechanical conditions, but the observations have not been validated in patients. Second, responses were not tested at a range of back-up rates and pressures and only two leak sizes were used. These choices were made in order to simplify data presentation, and to represent the range of settings commonly encountered clinically 10. The two leaks were selected to range below and above those measured in a previous clinical study 3, and ideal settings were used for NPPV based on recommendations in the literature 8, 9. Third, the effects of leak on ventilator triggering or cycling during spontaneous breathing were not tested, so the observations are most relevant to situations where controlled breathing predominates, such as in neuromuscular patients during sleep 3. Fourth, with advances in technology, newer versions of some of the ventilators tested might perform differently. Finally, one representative ventilator of each type was tested and that individual ventilators may differ was allowed for. The study was designed mainly to illustrate certain response patterns of ventilator modes in the presence of air leaking, and should not be construed as an efficacy comparison between specific ventilators in compensating for leaks. The present results should alert practitioners to the need for calibration and close monitoring of ventilators used for NPPV, and for careful selection of ventilator settings.
In conclusion, pressure-targeted modes maintain delivered VT in the presence of leaks better than volume-targeted modes. Hence, pressure-targeted ventilators are preferred over volume-targeted ventilators to provide more effective NPPV in patients with substantial air leaking. To best compensate for air leaks, pressure-targeted ventilators should have high and sustained maximal inspiratory flow capabilities (>3 L·s1), adjustable I:E ratios or other mechanisms to limit inspiratory duration so that inversion of the I:E ratio is avoided, and adjustable trigger sensitivities or algorithms to prevent autocycling.
| Acknowledgements |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Garpestad, J. Brennan, and N. S. Hill Noninvasive Ventilation for Critical Care Chest, August 1, 2007; 132(2): 711 - 720. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Maheshwari, D. Paioli, R. Rothaar, and N. S. Hill Utilization of Noninvasive Ventilation in Acute Care Hospitals: A Regional Survey Chest, May 1, 2006; 129(5): 1226 - 1233. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Miyoshi, Y. Fujino, A. Uchiyama, T. Mashimo, and M. Nishimura Effects of Gas Leak on Triggering Function, Humidification, and Inspiratory Oxygen Fraction During Noninvasive Positive Airway Pressure Ventilation Chest, November 1, 2005; 128(5): 3691 - 3698. [Abstract] [Full Text] [PDF] |
||||
![]() |
J M Tuggey and M W Elliott Randomised crossover study of pressure and volume non-invasive ventilation in chest wall deformity Thorax, October 1, 2005; 60(10): 859 - 864. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. F.M. Wijdicks, E. M. Manno, and S. R. Holets Ventilator self-cycling may falsely suggest patient effort during brain death determination Neurology, September 13, 2005; 65(5): 774 - 774. [Full Text] [PDF] |
||||
![]() |
M.W. Elliott The interface: crucial for successful noninvasive ventilation Eur. Respir. J., January 1, 2004; 23(1): 7 - 8. [Full Text] [PDF] |
||||
![]() |
H. Prigent, C. Samuel, B. Louis, M.-F. Abinun, F. Zerah-Lancner, M. Lejaille, J.-C. Raphael, and F. Lofaso Comparative Effects of Two Ventilatory Modes on Speech in Tracheostomized Patients with Neuromuscular Disease Am. J. Respir. Crit. Care Med., January 15, 2003; 167(2): 114 - 119. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Vitacca, L. Barbano, S. D'Anna, R. Porta, L. Bianchi, and N. Ambrosino Comparison of Five Bilevel Pressure Ventilators in Patients With Chronic Ventilatory Failure: A Physiologic Study Chest, December 1, 2002; 122(6): 2105 - 2114. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.P. Highcock, E. Morrish, S. Jamieson, J.M. Shneerson, and I.E. Smith An overnight comparison of two ventilators used in the treatment of chronic respiratory failure Eur. Respir. J., October 1, 2002; 20(4): 942 - 945. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Schonhofer and S. Sortor-Leger Equipment needs for noninvasive mechanical ventilation Eur. Respir. J., October 1, 2002; 20(4): 1029 - 1036. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |