|
|
||||||||
1 Dept of Neonatology (Charité), Humboldt University Berlin, Germany and 2 Portex Anaesthesia, Intensive Therapy and Respiratory Medicine Unit, Institute of Child Health, London, UK
CORRESPONDENCE: G. Schmalisch, Dept of Neonatology (Charité), Humboldt University Berlin, Schumannstraße 20/21, D-10098, Berlin, Germany. Fax: 49 3028025824
This work, was supported by the German Ministry for Research and Technology, project "Perinatal Lung" (grant 01 ZZ 9511) and Deutsche Forschungsgemeinschaft (Schm 1160/1-2).
The aim of this in vitro study was to compare the measurement accuracy of two currently available devices for measuring tidal breathing in infants.
A mechanical model pump was used to generate flow profiles which simulated those observed in infants. A range of flows was applied simultaneously to two different devices, namely the commercially available SensorMedics 2600 (SM 2600) and more recently developed, custom-made equipment based on the flow-through technique (FTT). Automatically derived values from both devices were compared with one another and with manual calculations of printouts of the same breaths.
There were no differences in the raw flow signal obtained from the two devices, nor between values calculated automatically or manually from the FTT. Similarly, the deviations between the FTT and SM 2600 were <3% for tidal volume, respiratory frequency and minute ventilation. However, when comparing either with manually calculated values or those derived automatically from the FTT, there was a systematic and highly significant underestimation of shape-dependent parameters, such as the time to peak tidal expiratory flow as a proportion of tidal expiratory time (tPTEF/tE), derived by the SM 2600. The lower the applied flow, the higher the observed deviations, the underestimation being up to 60% when flows simulating those observed in preterm neonates were applied.
These errors appear to result from differences in signal processing such as the algorithms used for breath detection and can only be detected if appropriate nonsinusoidal flow profiles representing those seen in infants are used to evaluate equipment.
In contrast to that used in adults, in vivo validation of equipment designed to measure respiratory function in infants is extremely difficult, due to the highly variable breathing pattern, limited period of sleep when recordings can be obtained and the potential influence of apparatus dead space on the recordings achieved 1. Consequently, greater reliance must be placed on in vitro evaluation in this age group. Unfortunately, in the past, such evaluations have often been performed with relatively large sinusoidal signals that do not necessarily represent the variety of flow signals that are observed during infant respiratory function tests (RFTs). Although several in vitro studies have been performed to assess precision of various infant RFT devices 25, a thorough assessment of measurement error has been difficult due to the wide range of signals encountered in this age group according to age, clinical condition and type of test. Thus while peak flow may be less than 30 mL·s1 during tidal breathing in a sick preterm infant, it may exceed 2 L·s1 during forced expiratory manoeuvres from raised lung volume in a healthy 18 month old. A European Respiratory Society/American Thoracic Society (ERS/ATS) task force is trying to address some of these problems by defining minimal performance criteria for equipment 6 and software 7.
During the last few years, several new techniques have been developed 8 and measurements are now more commonly performed with commercially available equipment than in the past. There is an urgent need for careful validation of such devices by in vitro measurements prior to any clinical or research applications. While the accuracy of volume measurements can easily be determined using calibration syringes 5, the accuracy of other tidal breathing parameters is more difficult to determine, since this requires a model pump with adjustable flow profiles if it is to provide a meaningful simulation of in vivo measurements 9. With increasing emphasis on the need for quality control and standardization of infant RFTs if these are to be used as reliable outcome measures in clinical or research studies, it is essential to compare the performance of different devices under conditions that closely resemble those likely to be encountered physiologically.
The aim of this study was to use a model pump that could deliver a variety of tidal volumes with variable flow profiles to compare the measurement accuracy of two currently available devices for measuring tidal breathing in infants.
| Methods |
|---|
|
|
|---|
|
|
Both devices allowed raw flow transducer signals to be recorded as an electrical voltage signal. For the SM 2600, this was achieved via a special service port of the device, whereas it was an integral part of the FTT system. There was comparable resolution of signal processing of the flow signals from both devices, with respect to both sampling rate (256 Hz for the SM 2600 and 200 Hz for the FTT) and A/D conversion (14 and 16 Bit A/D conversion, respectively). A warm up period of 30 min was allowed for both devices and calibration of the PNT, including the linearization procedure of the SM 2600, was performed prior to the measurements.
Protocol
For the evaluation, three tidal breathing parameters which depend only on the depth and frequency of the breathing cycle (tidal volume (VT), respiratory frequency (fR), minute ventilation (V'E)) were selected, together with four parameters which depend on the shape of the time based flow signal or tidal breathing flow-volume loop (TBFVL) (expiratory time (tE), time to peak tidal expiratory flow as a proportion of tidal expiratory time tPTEF/tE, volume to peak tidal expiratory flow as a proportion of tidal volume (VPTEF/VT) and flow when 25% of tidal volume remains in the lung (TEF25)).
Using the signals shown in figure 2
, attempts were made to validate the automatically calculated parameters by comparing them with those calculated manually from appropriate print outs. This was relatively simple for the FTT as the calibrated flow and volume signals could be printed out on a large enough scale to permit accurate manual evaluation (fig. 2
). Since this was not possible for the SM 2600 due to the relatively poor quality of the graphics, the tidal breathing parameters calculated by the SM 2600 were compared with those derived by manual evaluation of identical breaths that were measured and printed by the FTT. In addition, the uncalibrated raw flow signals of the SM 2600 were used to compare the values of timing parameters such as tE and tPTEF/tE that were calculated both manually and automatically by the SM 2600 (fig. 3
).
|
|
| (001) |
|
| (002) |
Statistical methods
Mean±sd were calculated for all ventilatory parameters and the comparability of parameters between the two devices was assessed using the method of Bland and Altman 15. The coefficient of variation (CV(%)=100x(sd/mean)) was used to describe the reproducibility of each parameter and the CVs for each device were compared by a rank test. Statistical evaluations were performed using the software STATGRAPHICS (Vers. 4.0, Manugistics Inc., Rockville, MD, USA). A level of statistical significance of p<0.05 was accepted.
| Results |
|---|
|
|
|---|
|
|
|
The variability of the calculated parameters was very low for VT, fR, V'E, tE and TEF25 (median CV<1.5%), although the CVs for the SM 2600 where generally significantly higher than those for the FTT (table 3
). Considerably larger CVs were found for tPTEF/tE, and VPTEF/VT for both devices.
|
| Discussion |
|---|
|
|
|---|
Comparison of both devices
The SM 2600 is currently the most frequently used and extensively investigated commercial device for RFT in newborns 1012, and hence was selected for this comparative study. In addition, during a recent assessment of the influence of apparatus deadspace on tidal breathing parameters in infants 1, marked discrepancies were identified in certain parameters over and above those that would be anticipated for physiological reasons, which warranted further investigation. Previous investigations have demonstrated that both the SM 2600 and the FTT can measure volume accurately (<3% error) over the range of 1060 mL 14, 18, and this was confirmed in the present study. However, there is no published information about the accuracy of shape dependent tidal breathing parameters such as tPTEF/tE which are widely used but much more difficult to evaluate.
The accuracy of the software in the FTT was relatively easy to validate, in that not only was there full access to the algorithms used, and the ability to export and inspect raw and calibrated signals such as ASCII files, but there was sufficient resolution in the graphics and printouts to enable accurate manual checks to be performed for the parameters that were automatically displayed by the device (figs. 2 and 3![]()
). However, since this was not possible with the SM 2600 due to the poor quality graphical printouts, detection of any flow-dependent measuring errors in tPTEF/tE and VPTEF/VT would have been much more difficult. The equipment and software of the SM 2600 (which is currently undergoing major review) is at least 10 yrs older than those of the FTT and significant technological progress in data processing and graphical display has occurred during this period. Nevertheless, the ability to access the raw flow signal from the SM 2600 via a special service port did allow some manual checks to be performed. Both devices have nearly the same flow resolution with regard to magnitude and time and, after numerical adjustments of the zero line and amplitude, it is evident that both raw flow signals are virtually the same (fig. 4
), Since there is no visible difference in the flow signals between the two devices, the observed differences in the displayed results must be the result of: 1) differences in the calibration factors; 2) numerical corrections (e.g. correction to body temperature, pressure, saturated (BTPS) conditions); 3) compensation and control of zero-line offset; or 4) the evaluation software.
Deviations in the measured parameters
The comparative in vitro measurements showed that, for parameters such as VT, fR and V'E, the observed differences between both devices were low over the whole volume range, and would not have any influence on the comparability of these parameters. By contrast, much larger deviations were found for tidal breathing parameters which depend on the shape of the flow and volume signal. Such parameters are much more influenced by signal processing (sample rate, analogue/digital conversion, filtering), noise, software and to a large extent, by the shape of the flow profile. The curvature of peak values on the flow signal is often relatively flat such that small disturbances in the flow signal can cause large deviations in tPTEF or VPTEF. This is reflected in the current study by the much higher CVs for tPTET and VPTEF than for VT, fR and V'E (table 3
).
Many of the observed deviations between the FTT and SM 2600 (tables 1 and 2![]()
) may have resulted from the use of different algorithms to detect the beginning of inspiration and expiration. As discussed in detail recently, this is an extremely complex area which can easily lead to errors, especially in the presence of low flows or noisy signals 19, 20. For the more recently developed FTT, the equipment and software had been designed to ensure optimal stabilization of the end-expiratory level 16 and breath detection, even in the presence of noisy signals 19. Since the SM 2600 is based on unpublished algorithms, it is difficult to identify the source of the discrepancies in tPTEF/tE and VPTEF/VT, between the two devices. The fact that there was close agreement in the values of tE calculated automatically, whether compared with those derived manually from the SM 2600 flow signal, or reported by the FTT, suggests that the error must lie in the way in which tPTEF is automatically calculated by the SM 2600.
In the current study, the deviations of tPTEF/tE and VPTEF/VT between the two devices were found to decrease with increasing magnitude of the flow and volume signal (table 2
). This dependence on flow for accuracy of displayed parameters often occurs in devices which use a dead band around the "zero flow" to stabilize the numerical integration of the signal (fig. 5
). In such a system, all flow values within the selected dead band are set to zero so that there is no volume drift when flow is in this range, as may occur for example during a long expiratory pause. As shown in figure 5
, such a dead band will have only a minimal influence on fR or tE but, as a result of the delay in recognizing start of expiration, may lead to significant underestimation of tPTEF, the magnitude of which will be inversely proportional to that of the flow signal. Similar effects will be obtained if thresholds of the volume signal are used to reduce the drift of the displayed volume. Further potential sources of measurement error when calculating tPTEF/tE and VPTEF/VT include failure to interpolate when identifying the exact moment of zero crossing between inspiration and expiration, and the presence of a zero-offset on the flow signal. The latter may arise from an undetected or poorly corrected off-set on the flow transducers or as the result of inappropriate correction of a volume drift.
|
Clinical implications
The SM 2600 represents the first generation of commercial equipment for infant respiratory function testing, and has made an important contribution to the clinical use of RFT in infancy. An unexpected finding from this study was that the measuring errors of the older device were not due to limitations of the sensors/transducers, but more to limitations in the software that was available at this time. Inability to perform simple checks on the accuracy of displayed parameters due to poor graphical resolution of the displayed and printed signals, together with the lack of suitable means of assessing performance formally, meant that any errors could easily go undetected for long periods of time. Since evaluations of this nature have rarely been performed, similar problems may well exist in other currently available systems, whether commercially available or custom built.
The results from this study suggest that equipment for infant RFTs needs to be carefully checked using a suitable model 9 before clinical application 4, 6, 18. When measuring tidal breathing, information about the accuracy of volume measurements alone is not sufficient, since the accuracy of tidal breathing parameters depends on measuring both the range and the shape of the flow signals. Extensive model measurements for infant RFT using different flow signal patterns have not been performed, and any previous evaluations have been based on relatively large sinusoidal signals that are rarely observed during clinical measurements.
The discrepancies in tPTEF/tE during both in vitro and in vivo studies were so large that values reported in the literature collected using different devices will clearly not be comparable. Furthermore, the relationship between the magnitude of error and absolute flows may seriously impair interpretation of the effects of disease, therapeutic interventions or growth and development on such parameters. There is therefore an urgent need for the user to have access to the algorithms and thresholds that are used in any specific device, and to understand how differences between devices may influence the results.
During recent years, there have been dramatic technological advances in both computer hardware and software. This has been accompanied by an increased commitment from both users and manufacturers to establish internationally approved standards in order to improve the confidence with which such equipment can be used 6, 7. Indeed, following recent discussions with the European Respiratory Society/American Thoracic Society task force on standardization of infant respiratory function tests, manufacturers have agreed that in future, details of relevant algorithms will be more readily available to the users. This would greatly facilitate meaningful evaluations and comparisons of different systems, and hopefully minimize the occurrence of the type of discrepancies identified in this study.
Received: June 30, 1999
Accepted July 26, 2000
This article has been cited by other articles:
![]() |
G. Schmalisch, B. Foitzik, R.R. Wauer, and J. Stocks Effect of apparatus dead space on breathing parameters in newborns: "flow-through" versus conventional techniques Eur. Respir. J., January 1, 2001; 17(1): 108 - 114. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Gappa, A.A. Colin, I. Goetz, and J. Stocks Passive respiratory mechanics: the occlusion techniques Eur. Respir. J., January 1, 2001; 17(1): 141 - 148. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |