Eur Respir J 1996; 9: 125-130
Copyright © ERS Journals Ltd 1996
Visual and different automatic scoring profiles of respiratory variables in the diagnosis of sleep apnoea-hypopnoea syndrome
O Carrasco,
JM Montserrat,
P Lloberes,
C Ascasco,
E Ballester,
C Fornas,
and
R Rodriguez-Roisin
The purpose of our study was to explore the diagnostic accuracy of different methods of scoring night time recording of respiratory variables (NTRRV) for the diagnosis of the sleep apnoea-hypopnoea syndrome (SAHS). Within a 2 week period, we performed a partially attended night time recording of respiratory variables and a full polysomnography (PSG) for reference in patients with suspected SAHS. Night time recording of respiratory variables was carried out using equipment which records, and continuously displays on a monitor, oximetry, airflow, chest and abdominal motion and body position. Night time recording of respiratory variables was scored manually and automatically, according to different combinations of the parameters described previously. Full polysomnography was performed in the Sleep Laboratory following conventional standards. Thirty six patients were studied. Visual analysis and different automatic scoring profiles of night time recording of respiratory variables were compared to full polysomnography in terms of agreement, sensitivity and specificity. Visual scoring of night time recording of respiratory variables gave the finest agreement-sensitivity-specificity relationship. Automatic scoring of nighttime recording of respiratory variables showed a trend to underestimate the apnoea-hypopnoea index (AHI) with respect to full polysomnography due mainly to underrecognition of hypopnoeas. Agreement-sensitivity-specificity relationships of automatic night time recording of respiratory variables with respect to full polysomnography varied depending on the automatic profile used. Some had a good agreement and sensitivity whilst others had a good specificity. These findings show that visual scoring of night time recording of respiratory variables is the most accurate method of analysis when compared to full polysomnography. The usefulness of the automatic methods of scoring of respiratory variables depends on the end-point chosen and is not reliable enough to be used in all situations. Night time recording of respiratory variables represents a real complement to conventional full polysomnography in clinical practice.
This article has been cited by other articles:

|
 |

|
 |
 
F. J. Ruiz-Lopez, B. Fernandez-Suarez, J. Guardiola-Martinez, I. Vergara-LaHuerta, J. Latour-Perez, and M. Lorenzo-Cruz
Quality Control of the Ambulatory Polygraphy Using Automatic Analysis
Chest,
January 1, 2009;
135(1):
194 - 200.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Campos-Rodriguez, N. Pena-Grinan, N. Reyes-Nunez, I. De la Cruz-Moron, J. Perez-Ronchel, F. De la Vega-Gallardo, and A. Fernandez-Palacin
Mortality in Obstructive Sleep Apnea-Hypopnea Patients Treated With Positive Airway Pressure
Chest,
August 1, 2005;
128(2):
624 - 633.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D Schlosshan and M W Elliott
Sleep * 3: Clinical presentation and diagnosis of the obstructive sleep apnoea hypopnoea syndrome
Thorax,
April 1, 2004;
59(4):
347 - 352.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. Netzer, A. H. Eliasson, C. Netzer, and D. A. Kristo
Overnight Pulse Oximetry for Sleep-Disordered Breathing in Adults : A Review
Chest,
August 1, 2001;
120(2):
625 - 633.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 1996 by the European Respiratory Society.
|