Three major problems are currently associated with airway stents: mucostasis, formation of granulation tissue, and migration. We wanted to determine whether these problems could be solved by a different stent design. Based on theoretical considerations of an idealized trachea, we developed a dynamic bifurcation stent made of silicone which incorporates horseshoe-shaped steel struts. A flexible posterior membrane enables dynamic compression during cough, whilst the clasps maintain the airway lumen in the face of external compression. The design of the stent cast was based upon computed tomographic (CT)-scan studies of the central airways. Its complex shape provides a smoother distribution of pressure on the mucosa; thereby, lowering the stimulus for granulation formation. The bronchial limbs saddle on the carina, preventing displacement. The mechanical behaviours of the new stent and two commercially available stents were compared in an ex-vivo model, utilizing freshly excised tracheae and new visualization techniques. Dynamic (artificial coughs) and static loads (simulating tumour compression or pleural pressures) were applied on excised human tracheae with different stents. Our dynamic stent preserved effective compression of the posterior membrane in response to cough, and also provided lumen stability against extrinsic compression. In comparison, the two commercially available stents did not provide both functions equally well. In conclusion, our newly designed dynamic bifurcation stent shows characteristics which should prove useful in avoiding problems currently associated with airway stents.
This article has been cited by other articles:

|
 |

|
 |
 
H.-S. Nam, S. W. Um, W.-J. Koh, G. Y. Suh, M. P. Chung, O. J. Kwon, J. Kim, and H. Kim
Clinical application of the Natural Y stent in the management of benign carinal stenosis.
Ann. Thorac. Surg.,
August 1, 2009;
88(2):
432 - 439.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Nakajima, Y. Sekine, M. Yasuda, K. Yasufuku, A. Iyoda, M. Suzuki, K. Shibuya, T. Iizasa, and T. Fujisawa
Long-Term Management of Polychondritis with Serial Tracheobronchial Stents
Ann. Thorac. Surg.,
June 1, 2006;
81(6):
e24 - e26.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Miwa, M. Mitsuoka, K. Tayama, N. Tomita, S. Takamori, A. Hayashi, and K. Shirouzu
Successful Airway Stenting Using Silicone Prosthesis for Esophagobronchial Fistula
Chest,
October 1, 2002;
122(4):
1485 - 1487.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J.F. Beamis, H.D. Becker, S. Cavaliere, H. Colt, J.P. Diaz-Jimenez, J.F. Dumon, E. Edell, K.L. Kovitz, H.N. Macha, A.C. Mehta, et al.
ERS/ATS statement on interventional pulmonology: Chairmen: C.T. Bolliger, P.N. Mathur
Eur. Respir. J.,
February 1, 2002;
19(2):
356 - 373.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Schmidt, H. Olze, A. C. Borges, M. John, U. Liebers, O. Kaschke, K. Haake, and C. Witt
Endotracheal balloon dilatation and stent implantation in benign stenoses
Ann. Thorac. Surg.,
May 1, 2001;
71(5):
1630 - 1634.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Nomori, H. Horio, Y. Imazu, and K. Suemasu
Double stenting for esophageal and tracheobronchial stenoses
Ann. Thorac. Surg.,
December 1, 2000;
70(6):
1803 - 1807.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Wassermann, F. Mathen, and H. E. Eckel
Malignant laryngotracheal obstruction: a way to treat serial stenoses of the upper airways
Ann. Thorac. Surg.,
October 1, 2000;
70(4):
1197 - 1201.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Noppen, M. Meysman, I. Claes, J. D'Haese, and W. Vincken
Screw-thread vs Dumon Endoprosthesis in the Management of Tracheal Stenosis
Chest,
February 1, 1999;
115(2):
532 - 535.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Wassermann, A. Koch, J. Muller-Ehmsen, M. Reuter, O. Michel, and H.-E. Eckel
CLINICAL AND LABORATORY EVALUATION OF A NEW THIN-WALLED SELF-EXPANDING TRACHEOBRONCHIAL SILICONE STENT: PROGRESS AND PITFALLS
J. Thorac. Cardiovasc. Surg.,
October 1, 1997;
114(4):
527 - 534.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
C. H. Marquette, E. Mensier, M.-C. Copin, A. Desmidt, L. Freitag, C. Witt, L. Petyt, and P. Ramon
Experimental Models of Tracheobronchial Stenoses: A Useful Tool for Evaluating Airway Stents
Ann. Thorac. Surg.,
September 1, 1995;
60(3):
651 - 656.
[Abstract]
[Full Text]
|
 |
|
Copyright © 1994 by the European Respiratory Society.
|