ERJ
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via ISI Web of Science (43)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gessler, T.
Right arrow Articles by Seeger, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gessler, T.
Right arrow Articles by Seeger, W.
European Respiratory Journal 17:14-19 (2001)
© 2001 ERS Journals Ltd


Ultrasonic versus jet nebulization of iloprost in severe pulmonary hypertension

T. Gessler1, T. Schmehl1, M.M. Hoeper2, F. Rose1, H.A. Ghofrani1, H. Olschewski1, F. Grimminger1 and W. Seeger1

1 Dept of Internal Medicine, Justus-Liebig-University of Giessen, Giessen, Germany, 2 Dept of Respiratory Medicine, Hannover Medical School, Hannover, Germany

CORRESPONDENCE: W. Seeger, Dept of Internal Medicine II, Justus-Liebig-University, Klinikstr. 36, Giessen, D-35392, Germany. Fax: 49 6419942359

Inhalation of iloprost, a stable prostacyclin analogue, is a promising perspective in the treatment of pulmonary hypertension. In initial clinical studies, a conventional jet nebulizer system was successfully used to decrease pulmonary vascular resistance and pressure, requiring however, up to twelve inhalations of 12–15 min per day. The aim of this study was to investigate if the application of an equal dose of iloprost at a drastically reduced duration of inhalation with the use of a more efficient ultrasonic nebulizer, leads to comparable haemodynamic effects, without escalation of side effects.

The physical features of the jet nebulizer system (Ilo-NebTM) and the ultrasonic nebulizer (Multisonic CompactTM) were characterized by laser diffractometry and a Tc99m-tracer technique. Mass median aerodynamic diameters were 3.2 µm for the jet and 3.9 µm for the ultrasonic nebulizer. Total output (mean±sd) was 60±7 µL·min–1 (jet) and 163±15 µL·min–1(ultrasonic), and efficiency of the devices was 39±3% (jet) and 86±5% (ultrasonic). Based on these data, a total inhalative dose of 2.8 µg iloprost was delivered by jet nebulization within 12 min and by ultrasonic nebulization within 4 min, in 18 patients with severe primary and secondary pulmonary hypertension (New York Heart Association class III and IV), in a randomized crossover design. Haemodynamics were assessed by right heart catheterization.

Inhalation with the ultrasonic device and jet nebulizer, reduced mean±sem pulmonary artery pressure from 54.3±2.1 to 47.1±2.0 and from 53.5±2.2 to 47.0±2.2 mmHg, respectively, and mean±sem pulmonary vascular resistance from 1073±109 to 804±87 and from 1069±125 to 810±83 dyn·s·cm–5, respectively. Both modes of aerosolization were well tolerated.

In conclusion, due to the markedly higher efficiency and output of the ultrasonic device, wastage of drug is largely avoided and the duration of inhalation can be shortened to one-third, with comparable haemodynamic effects and without enforcing side effects.

Keywords: iloprost, nebulization, pulmonary hypertension







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS