Eur Respir J 2000; 15: 338-341
Copyright © ERS Journals Ltd 2000
Inaccuracy of tidal volume delivered by home mechanical ventilators
F Lofaso,
R Fodil,
H Lorino,
K Leroux,
A Quintel,
A Leroy,
and
A Harf
Ideally, the inspired (tidal) volume (V(T)) provided by a volume-controlled ventilation device should not change when the pressure imposed on the ventilator varies. A bench study evaluation of V(T) versus pressure was performed on 10 commercially available devices. The difference between the desired V(T) and the observed V(T) reached 100 mL for some devices when inspiratory resistance was at its lowest, rising to 150 mL when inspiratory resistance was increased to obtain peak airway pressure of 60 cmH2O. The present data indicate that some home ventilators are inaccurate in delivering the preset tidal volume when the pressure imposed on the ventilator is increased to simulate high airway resistance.
This article has been cited by other articles:

|
 |

|
 |
 
B. Fauroux, K. Leroux, G. Desmarais, D. Isabey, A. Clement, F. Lofaso, and B. Louis
Performance of ventilators for noninvasive positive-pressure ventilation in children
Eur. Respir. J.,
June 1, 2008;
31(6):
1300 - 1307.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Gonzalez-Bermejo, V. Laplanche, F. E. Husseini, A. Duguet, J-P. Derenne, and T. Similowski
Evaluation of the user-friendliness of 11 home mechanical ventilators
Eur. Respir. J.,
June 1, 2006;
27(6):
1236 - 1243.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R Farre, D Navajas, E Prats, S Marti, R Guell, J M Montserrat, C Tebe, and J Escarrabill
Performance of mechanical ventilators at the patient's home: a multicentre quality control study
Thorax,
May 1, 2006;
61(5):
400 - 404.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Farre, S. J. Lloyd-Owen, N. Ambrosino, G. Donaldson, J. Escarrabill, B. Fauroux, D. Robert, B. Schoenhofer, A. Simonds, and J. A. Wedzicha
Quality control of equipment in home mechanical ventilation: a European survey
Eur. Respir. J.,
July 1, 2005;
26(1):
86 - 94.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Prigent, C. Samuel, B. Louis, M.-F. Abinun, F. Zerah-Lancner, M. Lejaille, J.-C. Raphael, and F. Lofaso
Comparative Effects of Two Ventilatory Modes on Speech in Tracheostomized Patients with Neuromuscular Disease
Am. J. Respir. Crit. Care Med.,
January 15, 2003;
167(2):
114 - 119.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Schonhofer and S. Sortor-Leger
Equipment needs for noninvasive mechanical ventilation
Eur. Respir. J.,
October 1, 2002;
20(4):
1029 - 1036.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M.W. Elliott, M. Confalonieri, and S. Nava
Where to perform noninvasive ventilation?
Eur. Respir. J.,
June 1, 2002;
19(6):
1159 - 1166.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. MEHTA and N. S. HILL
Noninvasive Ventilation
Am. J. Respir. Crit. Care Med.,
February 1, 2001;
163(2):
540 - 577.
[Full Text]
|
 |
|
Copyright © 2000 by the European Respiratory Society.
|