Eur Respir J 1998; 12: 759-763
Copyright © ERS Journals Ltd 1998
CPAP titration by an auto-CPAP device based on snoring detection: a clinical trial and economic considerations
M Berkani,
F Lofaso,
C Chouaid,
M Pia d'Ortho,
D Theret,
V Grillier-Lanoir,
A Harf,
and
B Housset
This study aimed to assess the ability of an auto-nasal continuous positive airway pressure (nCPAP) device (REM + auto; NPBFD, Nancy, France) to predict the optimal constant nCPAP level. The apnoea/hypopnoea detection facility of the auto-nCPAP device was deliberately disabled and nasal mask pressure vibration detection was the only mode of pressure setting. The auto-nCPAP device was tested on 10 previously untreated patients with obstructive sleep apnoea during a single night, with ambulatory polysomnography performed in a conventional hospital room; the efficacy of the fixed pressure determined by the auto-nCPAP device was assessed by an ambulatory full polysomnography 2 weeks after the initiation of treatment at home. The fixed nCPAP pressure was effective (apnoea/hypopnoea and arousal indices <10 events x h(-1)) in all but two of the 10 patients studied. When the fixed nCPAP pressure was increased by 2 cmH2O in these two patients, sleep and respiration were normalized. Since only 12 ambulatory polysomnographic recordings were used to determine the effective nasal continuous positive airway pressure level, and as the device restored normal breathing and sleep in all 10 patients, it was concluded that this method of nasal continuous positive airway pressure titration may improve cost-effectiveness and reduce waiting lists in sleep laboratories.
This article has been cited by other articles:

|
 |

|
 |
 
M. H. Sanders, J. M. Montserrat, R. Farre, and R. J. Givelber
Positive Pressure Therapy: A Perspective on Evidence-based Outcomes and Methods of Application
Proceedings of the ATS,
February 15, 2008;
5(2):
161 - 172.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Pevernagie, J. F. Masa, J. C. Meurice, R. Farre, O. Marrone, and D. Rodenstein
Treatment of obstructive sleep-disordered breathing with positive airway pressure systems
Eur. Respir. Rev.,
December 1, 2007;
16(106):
125 - 131.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. F. Masa, A. Jimenez, J. Duran, F. Capote, C. Monasterio, M. Mayos, J. Teran, L. Hernandez, F. Barbe, A. Maimo, et al.
Alternative Methods of Titrating Continuous Positive Airway Pressure: A Large Multicenter Study
Am. J. Respir. Crit. Care Med.,
December 1, 2004;
170(11):
1218 - 1224.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Noseda, C. Kempenaers, M. Kerkhofs, S. Braun, P. Linkowski, and E. Jann
Constant vs Auto-Continuous Positive Airway Pressure in Patients With Sleep Apnea Hypopnea Syndrome and a High Variability in Pressure Requirement
Chest,
July 1, 2004;
126(1):
31 - 37.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Lofaso, K. Leroux, M.A. Quera-Salva, G. Mroue, M.P. D'Ortho, D. Isabey, and B. Louis
Snoring detection during auto-nasal continuous positive airway pressure
Eur. Respir. J.,
January 1, 2002;
19(1):
108 - 112.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. I. Loube
Technologic Advances in the Treatment of Obstructive Sleep Apnea Syndrome
Chest,
November 1, 1999;
116(5):
1426 - 1433.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 1998 by the European Respiratory Society.
|