ERJ
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Permissions
Right arrowRequest Permissions
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gatnash, A.
Right arrow Articles by Connolly, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gatnash, A.
Right arrow Articles by Connolly, C.
Eur Respir J 1998; 12: 467-471
Copyright © ERS Journals Ltd 1998


Original Articles

A new method for measuring aerosol nebulizer output using radioactive tracers

AA Gatnash, ST Chandler, and CK Connolly

Reproducibility and comparability of bronchial challenge tests depends critically on accurate assessment of nebulizer output. Evaporation during nebulization means that simple weighing is inaccurate, overestimating the delivered dose of active ingredient. We wanted to quantify this effect in the context of intermittent nebulization, using a dosimeter as used in bronchial provocation tests. Output of three types of nebulizer, from the MEFAR dosimeter, was measured by radioactive tracer, using a standard solution of technetium-99m-pertechnetate (1.5 kBq x mL(-1)) in 4 mL of normal saline. The aerosol was impacted by suction onto a microfilter, and the radioactivity measured. Nebulizers were weighed before and after nebulization. Ratio of nebulized volume calculated from the radioactivity on the filter, to the total volume loss by weight, was expressed as nebulized ratio. The effect on output of two concentrations of methacholine, two tracers of different weights, and change in temperature, were assessed. Nebulized ratio varied between 44.1-71.6%. Results were more consistent within the same type of nebulizer than between different makes. Neither changes in molar concentration nor molecular weight affected nebulizer output or nebulized ratio. Mean nebulized ratio was 58.5%, showing that calibration by weighing, overestimates the delivered dose by a factor of approximately two. Measuring radioactivity eluted from a microfilter, onto which nebulized output had been impacted proved to be a satisfactory method of calibration.