ERJ
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Permissions
Right arrowRequest Permissions
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chiappini, F
Right arrow Articles by Pistelli, R
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chiappini, F
Right arrow Articles by Pistelli, R
Eur Respir J 1998; 11: 716-719
Copyright © ERS Journals Ltd 1998


Original Articles

Accuracy of a pulse oximeter in the measurement of the oxyhaemoglobin saturation

F Chiappini, L Fuso, and R Pistelli

The aim of the present study was to evaluate the accuracy, in comparison with a reference method, of the Nellcor N-20P pulse oximeter in the assessment of oxyhaemoglobin saturation (Sa,O2). Sa,O2 was monitored at rest by the Nellcor N-20P pulse oximeter in 100 subjects (82 males, mean age 68+/-12 yrs) consecutively enroled. At the same time, an arterial blood sample was collected for the measurement of Sa,O2, carboxyhaemoglobin, and methaemoglobin by an IL-282 Co-oximeter. A significant difference was found between Sa,O2 values measured with the two methods (t=11.78, p<0.05), but the two series of measurements were significantly correlated (r=0.97). Both the level and the limits of agreement between the two methods were satisfactory when the more appropriate Bland and Altman method was applied. Nevertheless, a lack of accuracy of the pulse oximeter was found, but only for Sa,O2 values <82% and >94%, as demonstrated by the Youden index. In conclusion, these data show that Nellcor N-20P is sufficiently reliable for the assessment and monitoring of oxyhaemoglobin saturation. The lack of accuracy does not seem clinically relevant since it is appreciable only for values at the extremes of the oxyhaemoglobin saturation range.


This article has been cited by other articles:


Home page
J. Appl. Physiol.Home page
K. A. Cope, M. T. Watson, W. M. Foster, S. S. Sehnert, and T. H. Risby
Effects of ventilation on the collection of exhaled breath in humans
J Appl Physiol, April 1, 2004; 96(4): 1371 - 1379.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1998 by the European Respiratory Society.